You can only observe an experiment in one flow channel at a time. Nevertheless, you can run subsequent experiments in all the channels: consecutive experiments can be set up for the different channels in the switchBUILD software by inserting a “new channel” element...
Eject the chip holder (without channel venting) and check if the chip was placed correctly in the chip holder. The chip was inserted correctly if you can easily read the “Dynamic Biosensors” label. Make sure to close the chip holder properly. This is the case, if you...
All DRX instruments can be equipped with different autosampler adaptors (standard tray, 96-well plate tray). To change the autosampler tray, please refer to the DRX manual or contact the Dynamic Biosensors support team at support@dynamic-biosensors.com. After exchange...
After using a flow channel on a switchSENSE® chip (i.e. after any type of liquid has entered the flow channel), it must be vented before it can be removed from the chip holder. This is easiest done during the chip ejection procedure: Go to „Chip“ Select „Eject Chip...
Both the Start and the Prime routines are default liquid handling procedures that are used to purge the DRX instrument with running buffers of interest. The Start routine should always be performed after a Clean&Sleep routine. In this thorough purging program, the...