During a measurement in dynamic mode, the DNA nanolevers are switched by applying an alternating voltage to the gold electrode, whereas in static mode they are kept in an upright orientation by a constant negative potential.
- The dynamic mode is usually used to investigate conformational changes.
- The static mode can be used for kinetic measurements as well as the investigation of bispecific binders, protein-protein interactions or enzymatic activity measurements.
Category: